STANDARD FEATURES
High Sensitivity-DC SQUIDS: Magnetic
Dipole Moment Noise of less than
10-12Am2 ( 10-9 EMU) RMS per root Hz for
7.6 cm access.
Low Helium Consumption - Approximately
900 days operation with 90 liters of helium.
No Helium Diffusion - All metal
construction eliminates cycling system to room temperature
to repump vacuum.
Heat Switch on Superconducting Shield
- Can change field with system filled with helium
in less than 10 minutes. About 0.2 liter of liquid per
change.
Sample Access - Vertical or Horizontal Orientation.
Standard Access diameter of 7.6
cm.
Pass Through Access for Long Cores.
Performance
is Independent of Liquid Helium Level .
Dynamic
Range - < 10-12 to > 10-4 Am2 analog in 4
ranges.
<
10-12 to > 10-3 Am2 digital flux counting on a single
range.
Compact
- Minimum length, maximum shielding.
Outside
diameter = 53 cm.
Overall
Length = 205 cm including cryocooler.
Cryocooler
Mounting - On opposite end from SQUIDS and electronics.
Field
Attenuation - 106 for transverse fields.
1011
for axial fields.
Microprocessor
Based Electronics -Digital and analog signal outputs,
with all range
and
filter settings front panel and computer controlled.
Digital
Flux Counting - Gives single scale with greater
than 180 db dynamic range.
Serial
RS-232 Data and Command - Link between SQUID electronics
and external
computer
to control electronics functions.
*
Available with one, two, or three measurement axes.
*
Manual and Computer Controlled Sample Handling and AF
Degaussing Systems Available.
|